CHD Master+ Micro hardness testers

Fully automatic, Economically, UP-TO-DATE
Including certified Vickers indenter, 5-megapixel camera system, Lenses: 4x, 20x, 40x, Export tool for XLS, Word, PDF

The CHD MASTER Variants

Q10 30 60 CHD MASTER

  • Attractive package price and short delivery times
  • Reduced to essentials: Vickers indenter and lenses included
  • Perfect for quick single-sample tests with CHD/NHD/SHD applications
  • Modular extension possible with sample image camera to variant “CHD Master+”
Q10 30 60 CHD MASTER +

  • Integrated 5 megapixels sample image camera
  • Single- and Multiple sample testing
  • Individual configurations: 3 model variants and load extension available
  • Expandable with innovative software modules

Test methods and force application

Test force variants

Q10 CHD MASTER+

0.25g
50 g
10 kg
*Test force extension (optional)
Vickers

Vickers

DIN EN ISO 6507, ASTM E-384, ASTM E92

  • HV 0.00025
  • HV 0.0005
  • HV 0.001
  • HV 0.002
  • HV 0.005
  • HV 0.01
  • HV 0.02
  • HV 0.025
  • HV 0.05
  • HV 0.1
  • HV 0.2
  • HV 0.3
  • HV 0.5
  • HV 1
  • HV 2
  • HV 3
  • HV 5
  • HV 10
Knoop

Knoop

DIN EN ISO 4545, ASTM E-384, ASTM E92

  • HK 0.00025
  • HK 0.0005
  • HK 0.001
  • HK 0.002
  • HK 0.003
  • HK 0.005
  • HK 0.01
  • HK 0.02
  • HK 0.025
  • HK 0.05
  • HK 0.1
  • HK 0.2
  • HK 0.3
  • HK 0.5
  • HK 1
  • HK 2
Brinell

Brinell

DIN EN ISO 6506, ASTM E-10

  • HBW 1/1
  • HBW 1/2.5
  • HBW 1/5
  • HBW 1/10

Q30 CHD MASTER+

0.25g
100 g
31.25 kg
*Test force extension (optional)
Vickers

Vickers

DIN EN ISO 6507, ASTM E-384, ASTM E92

  • HV 0.00025
  • HV 0.0005
  • HV 0.001
  • HV 0.002
  • HV 0.005
  • HV 0.01
  • HV 0.02
  • HV 0.025
  • HV 0.05
  • HV 0.1
  • HV 0.2
  • HV 0.3
  • HV 0.5
  • HV 1
  • HV 2
  • HV 3
  • HV 5
  • HV 10
  • HV 20
  • HV 30
Knoop

Knoop

DIN EN ISO 4545, ASTM E-384, ASTM E92

  • HK 0.00025
  • HK 0.0005
  • HK 0.001
  • HK 0.002
  • HK 0.003
  • HK 0.005
  • HK 0.01
  • HK 0.02
  • HK 0.025
  • HK 0.05
  • HK 0.1
  • HK 0.2
  • HK 0.3
  • HK 0.5
  • HK 1
  • HK 2
Brinell

Brinell

DIN EN ISO 6506, ASTM E-10

  • HBW 1/1
  • HBW 1/2.5
  • HBW 1/5
  • HBW 1/10
  • HBW 1/30
  • HBW 2.5/31.25

Q60 CHD MASTER+

0.25g
200 g
62.5 kg
*Test force extension (optional)
Vickers

Vickers

DIN EN ISO 6507, ASTM E-384, ASTM E92

  • HV 0.00025
  • HV 0.0005
  • HV 0.001
  • HV 0.002
  • HV 0.005
  • HV 0.01
  • HV 0.02
  • HV 0.025
  • HV 0.05
  • HV 0.1
  • HV 0.2
  • HV 0.3
  • HV 0.5
  • HV 1
  • HV 2
  • HV 3
  • HV 5
  • HV 10
  • HV 20
  • HV 30
  • HV 50
  • HV 60
Knoop

Knoop

DIN EN ISO 4545, ASTM E-384, ASTM E92

  • HK 0.00025
  • HK 0.0005
  • HK 0.001
  • HK 0.002
  • HK 0.003
  • HK 0.005
  • HK 0.01
  • HK 0.02
  • HK 0.025
  • HK 0.05
  • HK 0.1
  • HK 0.2
  • HK 0.3
  • HK 0.5
  • HK 1
  • HK 2
Brinell

Brinell

DIN EN ISO 6506, ASTM E-10

  • HBW 1/1
  • HBW 1/2.5
  • HBW 1/5
  • HBW 1/10
  • HBW 1/30
  • HBW 2.5/31.25
  • HBW 2.5/62.5
  • HBW 5/62.5 (> 30 HBW)

Convincing benefits

Ready to test

Ready to test

Vickers indenter and 3 lenses included in the basic package - no restrictions with test methods; all test load steps available and immediately useable.

Fully automatic test sequence

Fully automatic test sequence

With electronic load application and an enclosed circuit.

6-fold tool-change turret

6-fold tool-change turret

The 6-fold tool-change turret is standard and offers plenty of space for various test methods.

Optimum Stop

OPTIMUM STOP

Time-saving mode completes test series as soon as lower hardness limit has been undercut.

A simple start to fully-automated hardness testing

Large test space

Large test space

The sophisticated construction in anodized aluminium offers a large and well-arranged test space. The fully automatic XY-slide with a high- precision incremental measurement system can be equipped with an 1-fold sample holder. Beyond that, customer specific magazines can be managed and created in the software.

Revolutionary 3D concept

Revolutionary 3D concept

Intuitive, clearly structured and professional: Qpix Control2 heralds a new generation of hardness testing software developed to guarantee maximum user-friendliness.

Template function

Template function

  • Types of measurement: CHD, NHD, SHD, row measurement, rapid single measurements
  • Template function ideal for repeat tests / components
  • Pull test series into the desired position
  • The row starting point is automatically determined by the Auto-Snap function.
Protocol and data presentation

Protocol and data presentation

Wide range of structuration options for stored test item data. Adaptable content for export files and protocol content.

Modular expandable with sample image camera

Modular expandable with sample image camera

Modular upgrade with sample image camera for the CHD Master possible at any time – as “CHD MASTER +” ready for new applications.

Professional applications with the CHD MASTER+

Sample image camera

Sample image camera

Ultimate operating comfort provided by the USB 3.0 - 5 MP color camera system for capturing the whole sample and the perfect overview in test protocols.

Multiple sample testing and multi-bedding

Multiple sample testing and multi-bedding

With usage of Qness sample holders CHD Master+ variants enable tests on up to 8 embedded samples at a time – even in case several test pieces are embedded in one specimen.

2D/3D Mapping

2D/3D Mapping

Hardness distribution across partial segments or whole planes of test samples. Colourized representation in 2D and perspectively free twistable 3D on sample image.

Edge recognition

Edge recognition

The function of the Qness software module Edge recognition is the automatic adaption of test row start points to the sample edge when using project and sample templates. The module increases the machines automation and is the perfect add-on feature to the standardized AutoSnap function.

From CHD to welding samples

From CHD to welding samples

Further to quick and simple CHD/NHD/SHD testing which the machine is basically designed for, it is optional also possible to upgrade the machine with features for welding samples and visual test point patterns for extra-large work pieces.

Technical data

Q10/30/60 CHD MASTER+

 
Test force range Q10: 50 g - 10 kg (0,49 - 98,1 N)
Q30: 100 g - 31,25 kg (0,98 - 306,6 N)
Q60: 200 g - 62,5 kg (1,96 - 613,1 N)
Tool changer 6-fold, motorized
Sample image camera 5 megapixels
Types of measurement Multiple samples, CHD, NHD, RHD, Single piece-, row measurement
Test anvil/Cross slide motorized
Traverse path X/Y/Z X 150 / Y 150 / Z 145 mm
Weight of basic machine 58 kg
Data interface 1x USB3.0: Interface PC hardness tester
Repeatability X/Y slide +/- 3,5 µm
Camera system 5 megapixels
Power supply 230~1/N/PE, 110~1/N/PE
Max. power consumption ~ 200 W

Accessories and options
Includes lenses and indenter Lenses: 4x, 20x, 40x, DakkS- certified Vickers indenter
Test force extension (option) Q10: 0,25 g - 50 g (HV0,00025 - HV0,025 / ...)
Q30: 0,25 g - 100 g (HV0,00025 - HV0,5 / ...)
Q60: 0,25 g - 200 g (HV0,00025 - HV0,1 / ...)
Sample holder (option) 1-fold, 4-fold, 8-fold
Data connection plugins (option) Q-DAS, PCI, AMS, REMOTE

 

CHD Master+ Micro hardness testers

Fully automatic, Economically, UP-TO-DATE
Including certified Vickers indenter, 5-megapixel camera system, Lenses: 4x, 20x, 40x, Export tool for XLS, Word, PDF

This website uses cookies. By using our website, you also agree. Privacy Policy ›